OCT artifacts are image errors that can mimic disease, hide real pathology, or distort measurements.
Every clinician who uses OCT needs to recognize them, since treatment decisions often follow the scan. Most artifacts arise from the patient, the eye, or the software.
Why OCT Artifacts Matter
Retinal thickness maps guide anti-VEGF dosing, glaucoma monitoring, and surgical timing. A segmentation error can add or remove many micrometers, and it can trigger or delay treatment. Studies in glaucoma found artifacts in a large share of scans, and in macular scans the rate is also high (see OCT angiography).
Segmentation Errors
Software draws the boundaries of layers automatically, and errors occur in disease. Fluid, drusen, membranes, and atrophy confuse the algorithms, and the lines may jump to the wrong layer. Always inspect the B-scan, and correct the boundaries manually when measurements will guide treatment.
Fundus Explorer Pro
Photograph the retinal findings described here with the phone already in your pocket — 22 D optics and built-in illumination in one handheld unit.
From Choroida — the team behind this siteMotion and Blink OCT Artifacts
Eye movement causes discontinuities, doubled images, and stripes in en face maps, while blinks cause black bands across the scan. Tracking systems reduce these errors, though they cannot remove them in unsteady patients. Repeat the scan after a pause, and ask the patient to fixate.
Signal Strength and Media OCT Artifacts
Cataracts, corneal opacities, vitreous floaters, and small pupils reduce signal. A low-signal scan loses detail and may produce false thinning. Check the signal index, and consider a different device or wavelength when media are cloudy.
Mirror, Edge, and Shadow Artifacts
A mirror artifact occurs when the retina lies too close to the zero-delay line, and an inverted image appears. Ho and colleagues described these artifacts in high myopes and others. Edge cutoffs happen when the scan lies outside the window, and shadows from vessels or opacities hide deeper layers, so reposition the patient and rescan when these appear.
Preventing OCT Artifacts
- Dilate the pupil when the media permit
- Align the scan and adjust the focus before capture
- Check image quality before the patient leaves
- Compare with previous scans from the same device
Training the technician is the most effective measure (see anterior segment OCT).
An Example Case
A 70-year-old woman receives anti-VEGF injections for neovascular AMD, and her central thickness jumps from 280 to 410 micrometers on a follow-up scan. The macula looks dry on the B-scan, and the segmentation line has moved onto a drusen-related elevation. After manual correction, the thickness is 285 micrometers, and the clinician avoids an unnecessary injection.

Questions Clinicians Ask
Clinicians ask whether artifacts are common, and studies report them in a large share of scans. They also ask who should check the images, and the answer is the clinician who makes the decision, not the software. Because quality varies, build a routine for image review.
Pitfalls to Avoid
- Trusting thickness maps without opening the B-scan
- Accepting scans with a low signal index in a clinical decision
- Mixing measurements from different devices in a single trend
- Overreading a single outlier without repeating the scan
Looking Ahead
Better tracking and deep learning may reduce artifacts and flag poor scans automatically, and standardized quality indices may help clinicians to judge reliability. Until these tools mature, a careful look at each scan remains the best defense.
Practical Points
Teach staff to check alignment, focus, and signal before saving a scan, and then compare the new scan with prior ones to spot sudden changes. Also repeat the test when results do not fit the clinical picture.
Artifacts in Specific Diseases
In glaucoma, errors in nerve fiber layer thickness include segmentation faults and poor centration. Macular disease produces artifacts most often with fluid, drusen, and atrophy, while tilted or staphylomatous myopic eyes distort the layers and the scan geometry. Therefore, interpret thickness maps with the disease in mind.
Teaching Staff
Train technicians to recognize low-quality scans and to repeat them, and then set a minimum signal index in the clinic. Also keep a log of frequent artifacts and discuss them in staff meetings, so that everyone learns from errors. Artifacts also differ between devices, so learn how your instrument displays warnings and quality scores.
Because software updates change algorithms, review release notes when they arrive, and also ask the manufacturer about known artifacts for each scan protocol. Then share the answers with all staff who read the scans, and finally audit a sample of scans every few months and feed the results back to the team.


Document what you see
Two smartphone imaging tools built for everyday clinic use — one for the slit lamp, one for the fundus.
From Choroida — the team behind this siteReferences
- Asrani S, Essaid L, Alder BD, Santiago-Turla C. Artifacts in spectral-domain optical coherence tomography measurements in glaucoma. JAMA Ophthalmol. 2014;132:396-402.
- Ho J, Castro DP, Castro LC, et al. Clinical assessment of mirror artifacts in spectral-domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2010;51:3714-3720.
- Chhablani J, Krishnan T, Sethi V, Kozak I. Artifacts in optical coherence tomography. Saudi J Ophthalmol. 2014;28:81-87.
- American Academy of Ophthalmology. Basic and Clinical Science Course, Section 12: Retina and Vitreous. AAO; current edition.
Test yourself
A few questions straight from this article.
-
What is an OCT artifact in clinical retinal imaging?
The article defines OCT artifacts as image errors that can mimic disease, hide real pathology, or distort measurements. Most arise from the patient, the eye, or the software. -
Why do segmentation errors occur on OCT in eyes with macular disease?
Software draws layer boundaries automatically, and in disease, fluid, drusen, membranes, and atrophy confuse the algorithms. The lines may then jump to the wrong layer. -
How should OCT segmentation errors be handled when thickness measurements will guide treatment?
The article advises always inspecting the B-scan and correcting boundaries manually when measurements will guide treatment. -
What appears on OCT scans when a patient blinks during acquisition?
Blinks cause black bands across the scan, whereas eye movement causes discontinuities, doubled images, and stripes in en face maps. -
What is the recommended response to motion artifact on OCT in an unsteady patient?
Tracking reduces motion errors but cannot remove them in unsteady patients. The article advises repeating the scan after a pause and asking the patient to fixate. -
What effect can a low-signal OCT scan have on retinal measurements?
Cataracts, corneal opacities, floaters, and small pupils reduce signal. A low-signal scan loses detail and may produce false thinning, so check the signal index. -
What causes a mirror artifact on OCT, producing an inverted image?
A mirror artifact occurs when the retina lies too close to the zero-delay line, and an inverted image appears. It has been described in high myopes among others. -
In the article's example, a patient with neovascular AMD on anti-VEGF has central thickness jump from 280 to 410 micrometers. What explained it?
The macula looked dry on the B-scan, and the segmentation line had moved onto a drusen-related elevation. After manual correction thickness was 285 micrometers, avoiding an unnecessary injection. -
Which OCT trend-monitoring practice does the article list as a pitfall to avoid?
Among the pitfalls listed is mixing measurements from different devices in a single trend. The article recommends comparing with prior scans from the same device.