OCT artifacts are image errors that can mimic disease, hide real pathology, or distort measurements.

Every clinician who uses OCT needs to recognize them, since treatment decisions often follow the scan. Most artifacts arise from the patient, the eye, or the software.


Why OCT Artifacts Matter

Retinal thickness maps guide anti-VEGF dosing, glaucoma monitoring, and surgical timing. A segmentation error can add or remove many micrometers, and it can trigger or delay treatment. Studies in glaucoma found artifacts in a large share of scans, and in macular scans the rate is also high (see OCT angiography).

Segmentation Errors

Software draws the boundaries of layers automatically, and errors occur in disease. Fluid, drusen, membranes, and atrophy confuse the algorithms, and the lines may jump to the wrong layer. Always inspect the B-scan, and correct the boundaries manually when measurements will guide treatment.

Test your knowledge 9 questions from this article
Choroida · Fundus imaging

Fundus Explorer Pro

Photograph the retinal findings described here with the phone already in your pocket — 22 D optics and built-in illumination in one handheld unit.

From Choroida — the team behind this site

Motion and Blink OCT Artifacts

Eye movement causes discontinuities, doubled images, and stripes in en face maps, while blinks cause black bands across the scan. Tracking systems reduce these errors, though they cannot remove them in unsteady patients. Repeat the scan after a pause, and ask the patient to fixate.

Signal Strength and Media OCT Artifacts

Cataracts, corneal opacities, vitreous floaters, and small pupils reduce signal. A low-signal scan loses detail and may produce false thinning. Check the signal index, and consider a different device or wavelength when media are cloudy.

Mirror, Edge, and Shadow Artifacts

A mirror artifact occurs when the retina lies too close to the zero-delay line, and an inverted image appears. Ho and colleagues described these artifacts in high myopes and others. Edge cutoffs happen when the scan lies outside the window, and shadows from vessels or opacities hide deeper layers, so reposition the patient and rescan when these appear.

Preventing OCT Artifacts

  • Dilate the pupil when the media permit
  • Align the scan and adjust the focus before capture
  • Check image quality before the patient leaves
  • Compare with previous scans from the same device

Training the technician is the most effective measure (see anterior segment OCT).

An Example Case

A 70-year-old woman receives anti-VEGF injections for neovascular AMD, and her central thickness jumps from 280 to 410 micrometers on a follow-up scan. The macula looks dry on the B-scan, and the segmentation line has moved onto a drusen-related elevation. After manual correction, the thickness is 285 micrometers, and the clinician avoids an unnecessary injection.

OCT Artifacts: How to Recognize and Avoid Them

Questions Clinicians Ask

Clinicians ask whether artifacts are common, and studies report them in a large share of scans. They also ask who should check the images, and the answer is the clinician who makes the decision, not the software. Because quality varies, build a routine for image review.

Pitfalls to Avoid

  • Trusting thickness maps without opening the B-scan
  • Accepting scans with a low signal index in a clinical decision
  • Mixing measurements from different devices in a single trend
  • Overreading a single outlier without repeating the scan

Looking Ahead

Better tracking and deep learning may reduce artifacts and flag poor scans automatically, and standardized quality indices may help clinicians to judge reliability. Until these tools mature, a careful look at each scan remains the best defense.

Practical Points

Teach staff to check alignment, focus, and signal before saving a scan, and then compare the new scan with prior ones to spot sudden changes. Also repeat the test when results do not fit the clinical picture.

Artifacts in Specific Diseases

In glaucoma, errors in nerve fiber layer thickness include segmentation faults and poor centration. Macular disease produces artifacts most often with fluid, drusen, and atrophy, while tilted or staphylomatous myopic eyes distort the layers and the scan geometry. Therefore, interpret thickness maps with the disease in mind.

Teaching Staff

Train technicians to recognize low-quality scans and to repeat them, and then set a minimum signal index in the clinic. Also keep a log of frequent artifacts and discuss them in staff meetings, so that everyone learns from errors. Artifacts also differ between devices, so learn how your instrument displays warnings and quality scores.

Because software updates change algorithms, review release notes when they arrive, and also ask the manufacturer about known artifacts for each scan protocol. Then share the answers with all staff who read the scans, and finally audit a sample of scans every few months and feed the results back to the team.


All-fit smartphone adapter on a slit lampFundus Explorer Pro smartphone fundus camera
Choroida · Clinical imaging

Document what you see

Two smartphone imaging tools built for everyday clinic use — one for the slit lamp, one for the fundus.

From Choroida — the team behind this site

References

  1. Asrani S, Essaid L, Alder BD, Santiago-Turla C. Artifacts in spectral-domain optical coherence tomography measurements in glaucoma. JAMA Ophthalmol. 2014;132:396-402.
  2. Ho J, Castro DP, Castro LC, et al. Clinical assessment of mirror artifacts in spectral-domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2010;51:3714-3720.
  3. Chhablani J, Krishnan T, Sethi V, Kozak I. Artifacts in optical coherence tomography. Saudi J Ophthalmol. 2014;28:81-87.
  4. American Academy of Ophthalmology. Basic and Clinical Science Course, Section 12: Retina and Vitreous. AAO; current edition.

Test yourself

A few questions straight from this article.

1 / 9 0 correct
  1. What is an OCT artifact in clinical retinal imaging?

Ahmed Ophthalmologist

Writes and reviews the clinical articles on Educate. Every post is checked against peer-reviewed sources before it goes live.